“BURN OUT” is always displayed even though a thermocouple is connected.
Remove the thermocouple from the relevant channel and try short-circuiting between the + and – terminals with a short conductor. If the display shows room temperature (*) after short-circuiting, the thermocouple is disconnected. If “BURN OUT” is still displayed at all times, the main unit may be malfunctioning.
Please contact Graphtec distributor.
* The above is the case when pressing [MENU] > “OTHER” > “TEMP. Setting” > ”Room Temp.” is set to “Internal”.
When “Room Temp.” is set to “External”, approximately 0°C is displayed.