Instrument products > Handheld Measuring Instruments

Handheld Measuring Instruments

Handheld Measuring Instruments
Handheld Measuring Instruments

Digital multimeter

MM-300 Discontinued Product

High-precision, compact digital multimeter

  • 4,000-count large-format LCD
  • High-precision, high-voltage, multifunctional pocket-sized model
  • Allows measurement of maximum/minimum values and deviation

General specifications

Display
4,000-count LCD, 14-mm character height
Sampling
3 times/second
Range switching
Auto/manual ranging
Power source
One 3-V battery
Dimensions/weight
118 (H) x 78 (W) x 16 (D) mm/110 g
Accessory provided
Battery (3 V) installed

0.5% class DC/VCE

MM-300

  Range Resolution Accuracy Input resistance
Direct-current voltage 400mV
4.0V
40V
100μV
1mV
10mV
±0.5%rdg±3dgt 100MΩ or higher
11MΩ (approx.)
10MΩ (approx.)
400V/600V 100mV/1V ±1.0%rdg±3dgt
Alternating-current voltage 4.0V
40V/400V
600V
1mV
10mV/100mV
1V
±1.5%rdg±5dgt 11MΩ (approx.)
10MΩ (approx.)
Resistance Range Resolution Accuracy Test current
400Ω
4.0kΩ
40kΩ
400kΩ
4.0MΩ
40MΩ
0.1Ω

10Ω
100Ω
1kΩ
10kΩ
±1.5%rdg±4dgt 0.2mA or smaller
50μA or smaller
5μA or smaller
0.5μA or smaller
±1.0%rdg±3dgt
±3.0%rdg±3dgt 50nA
±7.0%rdg±3dgt
Other functions Frequency, duty ratio, continuity test, diode test, electrostatic capacity, maximum/minimum values, deviation value, display hold, auto power save

Digital clamp meters

CM-205

Optimal for switchboard measurement

  • One key switch enables all operations
  • Highly twist-resistant; beak-shaped clamp head provides easy attachment
  • The auto power-off function automatically cuts off the power source after 10 minutes.
  • The auto-ranging function eliminates annoying range switching.
  • Display hold function
  • Carrying case and spare fuse provided

General specifications

Display
4,000-count LCD, maximum display: 4,301
Sampling
2.5 times/second
Range switching
Auto ranging
Diameter of conductor
under measurement
27 mm maximum
 
Power source
One 3-V battery
Dimensions/weight
150 (H) x 52 (W) x 20 (D) mm/110 g
Accessory provided
Battery (3 V) installed, spare fuse, carrying case

AC 200ACE

CM-205

  Range Accuracy
Alternating current 40A/200A ±1.8%rdg±5dgt
Other functions Display hold, auto power off

CM-207

Large-format clamp tester that provides a full range of measurement functions including frequency measurement

  • Both auto-ranging and range hold (where the range is fixed to a specified range) measurement enabled
  • Dust-proof and drip-proof case construction protects the unit’s interior from dust and drips

General specifications

Display
4,000-count LCD, maximum display: 4,999
Sampling
3 times/second
Range switching
Auto ranging
Diameter of conductor
under measurement
55 mm maximum
 
Power source
One 9-V battery
Dimensions/weight
250 (H) x 92 (W) x 39 (D) mm/470 g
Accessory provided
Test leads, battery (9 V), carrying case

AC 2000ATrue RMSCE

CM-207

  Measurement range Accuracy
(50Hz-60Hz) (40Hz-1kHz)
A a.c.
Hz·%
Alternating
current
400A (0-600A) ±1.5%rdg±5dgt ±3.0%rdg±5dgt
2000A (0-1000A) ±1.5%rdg±5dgt ±3.0%rdg±5dgt
(1001A-) ±3.0%rdg±5dgt
Frequency 5Hz-4,999kHz ±0.2%rdg±2dgt
Duty ratio Not regulated
    (40Hz-400Hz) (401Hz-1kHz)
V a.c./V d.c.
Hz·%
Alternating
current voltage
4/40/400/600V ±1.5%rdg±5dgt ±3.0%rdg±5dgt
Direct-current voltage 400mV/4/40/400/600V ±1.2%rdg±3dgt
Frequency 1Hz-100kHz ±0.2%rdg±2dgt
Duty ratio 0.1%-99.9% ±2%rdg±2dgt
Resistance 400Ω ±.5%rdg±5dgt
4k/40k/400kΩ ±1.0%rdg±3dgt
4MΩ ±3.0%rdg±4dgt
40MΩ ±5.0%rdg±7dgt
Other functions continuity test, diode test, electrostatic capacity, maximum/minimum values, deviation value, range hold, display hold, auto power save

CM-208

All-purpose clamp meter that is optimal for electrical facility and vehicle maintenance/inspection

  • Allows continuity tests, diode tests, and measurement of electrostatic capacity
  • Measurement of maximum/minimum values, peak hold, deviation measurement
  • Dust-proof and drip-proof case construction protects the unit’s interior from dust and drips.

General specifications

Display
4,000-count LCD, maximum display: 9,999
Sampling
3 times/second
Range switching
Auto ranging
Diameter of conductor
under measurement
35 mm maximum
 
Power source
One 9-V battery
Dimensions/weight
193 (H) x 60 (W) x 34.5 (D) mm/300 g
Accessory provided
Test leads, battery (9 V), carrying case

AC/DC 1000ATrue RMSCE

CM-208

  Measurement range Accuracy
A a.c.
Hz·%
Alternating
current
400A ±1.5%rdg±5dgt
1000A (400-600A) ±1.5%rdg±5dgt
(601-1000A) ±3.0%rdg±5dgt
Direc
current
400A ±1.5%rdg±3dgt
1000A (400-600A) ±1.5%rdg±3dgt
(601-1000A) ±3.0%rdg±3dgt
Frequency 1Hz-1kHz ±0.2%rdg±2dgt
V a.c./V d.c.
Hz·%
Alternating-current
voltage
4/40/400/600V ±1.0%rdg±5dgt (40-400Hz)
Direct-current
voltage
400mV/4/40/400/600V ±1.0%rdg±3dgt
Frequency 1Hz-49.99kHz ±0.2%rdg±2dgt
Duty ratio 10%-90% ±2.0%rdg±2dgt
Resistance 400Ω ±1.5%rdg±5dgt
4k/40k/400kΩ ±1.0%rdg±3dgt
4MΩ ±4.0%rdg±5dgt
40MΩ ±6.0%rdg±5dgt
Other functions Continuity test, diode test, electrostatic capacity, maximum/minimum values, peak hold, deviation value, display hold, auto power save

CM-209

Enables checking of the frequency and duty ratio while alternating current/voltage measurement is being performed

  • The respective maximum/minimum values can be measured while alternating current, voltage, and resistance measurements are being performed.
  • Dust-proof and drip-proof case construction protects the unit’s interior from dust and drips.

General specifications

Display
4,000-count LCD, maximum display: 5,120
Sampling
3 times/second
Range switching
Auto ranging
Diameter of conductor
under measurement
55 mm maximum
 
Power source
One 9-V battery
Dimensions/weight
250 (H) x 92 (W) x 39 (D) mm/500 g
Accessory provided
Test leads, battery (9 V), carrying case

AC/DC 2000ATrue RMSCE

CM-209

  Measurement range Accuracy
  (50Hz-60Hz) (40Hz-1kHz)
A a.c.
Hz·%
Alternating
current
400A ±1.5%rdg±5dgt ±3.0%rdg±5dgt
2000A (0-1000A) ±1.5%rdg±5dgt ±3.0%rdg±5dgt
(1001A-) ±3.0%rdg±5dgt
Direc current 400/2000A ±1.5%rdg±5dgt
Frequency 1Hz-4.999kHz ±0.2%rdg±2dgt
    (40Hz-400Hz) (401Hz-1kHz)
V a.c./V d.c.
Hz·%
Alternating-current
voltage
4/40/400/600V ±1.5%rdg±5dgt ±3.0%rdg±5dgt
Direct-current
voltage
400mV/4/40/400/600V ±1.2%rdg±3dgt
Frequency 1Hz-100kHz ±0.2%rdg±2dgt
Duty ratio 0.1%-99.9% ±2.0%rdg±2dgt
Resistance 400Ω ±1.5%rdg±5dgt
4k/40k/400kΩ ±1.0%rdg±3dgt
4MΩ ±3.0%rdg±4dgt
40MΩ ±5.0%rdg±7dgt
Other functions Continuity test, diode test, electrostatic capacity, maximum/minimum values, deviation value, peak hold, display hold, auto power save

CM-211

A clamp meter of extraordinary versatility, with a four-item display of multiple measurement elements

  • The following parameters can be displayed concurrently while alternating current/voltage measurements are being performed.
    • Frequency/duty ratio/pulse width
    • Maximum value/minimum value/mean value and their duration
  • Output terminal provided for connection to an oscilloscope or recorder
  • Dust-proof and drip-proof case construction protects the unit’s interior from dust and drips

AC/DC 2000ATrue RMSCE

CM-211

General specifications

Display
4,000-count LCD, maximum display: 9,999
Sampling
1,000 times/second
Range switching
Auto/manual ranging
Diameter of conductor
under measurement
55 mm maximum
 
Power source
One 9-V battery or AC adapter (sold separately)
Dimensions/weight
250 (H) x 92 (W) x 39 (D) mm/500 g
Accessory provided
Test leads, battery (9 V), carrying case
Option
Temperature probe RIC-110

  Measurement range Accuracy
A d.c. + V d.c. Direct current 400/2000A ±1.5%rdg±3dgt
Direct-current voltage 40/400/600V ±1.0%rdg±3dgt
A a.c. + Hz
%/ms
Alternating current 400/2000A ±1.5%rdg±3dgt
Frequency 10Hz-25kHz ±0.2%rdg±2dgt
Duty ratio 1%-99% ±5.0%rdg±2dgt
Pulse width 0.1ms-100ms ±5.0%rdg±2dgt
V a.c. + Hz
%/ms
Alternating-current
voltage
40/400/600V ±1.5%rdg±3dgt
Frequency 10Hz-25kHz ±0.2%rdg±2dgt
Duty ratio 1%-99% ±5.0%rdg±2dgt
Pulse width 0.1ms-100ms ±5.0%rdg±2dgt
Resistance 4k/40kΩ ±1.0%rdg±2dgt
Other functions Temperature*1, continuity test, diode test, maximum/minimum values, mean value, peak hold, deviation value, output terminal, display hold, auto power save

*1 The temperature probe (RIC-110), sold separately, is required.

CM-114

An AC/DC clamp that displays both alternating current and voltage: optimal for vehicle and machine maintenance

  • Two electrical elements are displayed concurrently on the LCD while they are being measured.
  • Output terminal provided for connection to an oscilloscope or recorder
  • Dust-proof and drip-proof case
    construction protects the unit’s interior from dust and drips

General specifications

Display
4,000-count 2-item LCD, maximum display: 9,999
Sampling
15 times/second; 3.75 times/second when both direct current and direct-current voltage measurements are being performed in the normal mode
Range switching
Auto ranging
Diameter of conductor
under measurement
36 mm maximum
 
Power source
One 9-V battery or AC adapter (sold separately)
Dimensions/weight
200 (H) x 64 (W) x 33 (D) mm/310 g
Accessory provided
Test leads, battery (9 V), carrying case

AC/DC 1000ATrue RMS

CM-114

    Measurement range Accuracy
A d.c. + V d.c. Direct current 400A ±1.5%rdg±3dgt
1000A (400-600A) ±1.5%rdg±1dgt
(601-1000A) ±3.0%rdg±3dgt
Direct-current voltage 40V ±1.0%rdg±3dgt
250V ±1.0%rdg±1dgt
    (50Hz-60Hz) (40Hz-400Hz)
A a.c. + Hz
%/ms
Alternating current 400A ±1.5%rdg±3dgt ±2.5%rdg±3dgt
1000A (400-600A) ±1.5%rdg±1dgt ±2.5%rdg±1dgt
(601-1000A) ±3.0%rdg±3dgt ±4.0%rdg±3dgt
Frequency 10.0Hz-4000Hz ±0.5%rdg±1dgt
Duty ratio 0.0%-99.9% ±3.0%rdg±3dgt
Pulse width 1.0ms-999.9ms ±0.5%rdg±3dgt
V a.c. + Hz
%/ms
Alternating-current
voltage
40/250V (50Hz-60Hz) ±1.0%rdg±3dgt
(40Hz-400Hz) ±1.5%rdg±3dgt
Frequency 10.0Hz-4000Hz ±0.5%rdg±1dgt
Duty ratio 0.0%-99.9% ±3.0%rdg±3dgt
Pulse width 1.0ms-999.9ms ±0.5%rdg±3dgt
Resistance 4kΩ ±1.0%rdg±3dgt
20kΩ ±1.0%rdg±1dgt
Temperature*1 -30°C-150°C ±1.0%rdg±3dgt
Other functions Peak/maximum/minimum values, deviation value, display hold, output terminal

*1 The temperature probe (RIC-110), sold separately, is required.

Options

  • Temperature probe RIC-110

CM-113 Discontinued Product

Leakage current clamp meter equipped with a comparator

  • Beeper sounds when the specified comparator value (1 mA to 20 mA) is exceeded
  • Can be connected to a recorder for data recording
  • Low-pass filter: Cuts high frequencies of 150 Hz or higher

General specifications

Display
3,200-count LCD, maximum display: 3,200
Sampling
2 times/second
Range switching
Manual ranging
Diameter of conductor
under measurement
30 mm maximum
 
Power source
Two 1.5-V batteries
Dimensions/weight
212 (H) x 42 (W) x 30 (D) mm/220 g
Accessory provided
Two batteries (1.5 V), carrying case

AC 60A

CM-113

  Range Accuracy
Alternating current 30/300mA ±1.0%rdg±5dgt
60A (0-50A) ±1.0%rdg±5dgt
(50.01A-60.0A) ±5.0%rdg±5dgt
Other functions Analog output, maximum/minimum values, comparator (up to 20 mA), highfrequency cut-off, display hold, auto power off

CM-112 Discontinued Product

Connects to a measuring instrument to enable current measurement

  • When connected to a measuring instrument capable of measuring 400 mV AC/DC, the CM-112 enables measurement up to 400 AAC/ADC.
  • Specification: 0 to 400 AAC/ADC

AC/DC 400A

CM-112

Insulation resistance meters

ZM-510/520

  • New JIS Standard series carrying the CE Marking
  • Dust-proof and drip-proof construction; built-in timer
  • Hot lead detection; electrostatic discharge function

General specifications

Power source
Six 1.5-V R6P batteries
Dimensions/weight
169 (H) x 148 (W) x 50 (D) mm/570 g
Accessory provided
Test leads, six batteries (1.5 V), alligator clip (black lead only), test prod (with test pin/red lead only), spare fuse

CE

ZM-510ZM-520


  ZM-510 ZM-520
Standard
measurement
range
Insulation resistance Rated
voltage/resistance
50V/100MΩ 25V/5MΩ
125V/20MΩ 500V/100MΩ
  1000V/2000MΩ
Effective
measurement range
0.005-10MΩ 0.01-50MΩ
0.01-20MΩ 0.05-100MΩ
  1-2000MΩ
Center scale 0.2MΩ 1MΩ
0.5MΩ 2MΩ
  50MΩ
Output voltage on open circuit Within 1.3 times the rated voltage
Rated measurement current 1 mA to 1.2 mA
Short current Within 2 mA
  0 to 300 V 0 to 600 V
Precision Insulation resistance First effective measurement range: ±5%
Second effective measurement range: ±10%
Scale outside the above ranges: Within 0.5 mm of the scale length
Alternating-current voltage ±3% of the maximum scale value
Other functions Timer function (continuous measurement for approximately three minutes), battery check (through indication on the scale face)