Instrument products > midi LOGGER

GL800

Technical Specs & Details

Specifications

GL800 Main Unit Specifications

Item Description
Number of analog input terminal units 1unit (20 ch) or Extension unit (max. 200 ch)
Maximum 10 units or 500 ch when using with a PC
Sampling interval *1 100 ms (10 ch) to 1 h
Trigger function Type Start (Data capture starts when a trigger is generated)
Stop (Data capture stops when a trigger is generated)
Condition Start Off, Level, Alarm, Scheduled time, External
Stop Off, Level, Alarm, Scheduled time, Elasped time, External
Alarm function Type Analog, Logic, Pulse (AND and OR operation available)
Condition Analog H, L, Window In, Window Out
Logic H, L at each channel
Pulse H, L, Window In, Window Out
Pulse / Logic input *2 *5 Selectable between Pulse and Logic, Number of channels : 4
Pulse input range Count mode 50 C, 500 C, 5 kC, 50 kC, 500 kC, 5 MC, 50 MC, 500 MC/F.S.
(max. input 50kHz)
Inst. Mode 50 C, 500 C, 5 kC, 50 kC, 500 kC, 5 MC, 50 MC, 500 MC/F.S.
(max. input 50kHz)
RPM mode 50 rpm, 500 rpm, 5 krpm, 50 krpm, 500 krpm, 5 Mrpm,
50 Mrpm, 500 Mrpm/F.S. (max. input 50kHz)
Alarm output *5 No. of channels 4
Output format Open collector output (5V pull-up resistance 10 k ohm),5 to 24V (100 mA or less)
Output conditions Level judgement, Window judgement, Logic pattern judgement, Pulse judgement
External trigger input *2 *5 1 ch
Interface to PC Ethernet (10BASE-T / 100BASE-TX), USB (Compatible with high speed)
Data storage function Measured data Internal flash memory or USB memory directly
Other Setting conditions and Screen hard copy can be saved into the internal memory
or the USB memory
Internal storage device Internal flash memory : 12 Mbyte
USB memory slot (Full speed) Provided as standard
Calculation function Statistics calculation Average, Peak, maximum, Minimum, RMS
Number of calculations 2 calculations can be set simultaneously
Searching function Searching the necessary point from captured data. Type : Level, Alarm, Logic, Pulse
Display Size 5.7 inch TFT color LCD
Displayed items Waveforms + Digital values, Waveforms only, Digital values only
Operating environment Temperature : 0 to 45 °C (15 to 40 °C when operating by battery), Humidity : 5 to 85 % R.H.
Power supply AC adapter : 100 to 240V AC, 50/60 Hz
DC input : 8 to 24 V DC *3
Battery pack : Max. 2 batteries installable, 9 hours operation
(when using under Graphtec specified condition) *3
Power consumption 28 VA or lower (when operating with AC power)
External dimensions (W x D x H) (approx.) 232 mm x 152 mm x 50 mm
Weight (approx.) 990 g including a 20 channel input terminal unit *4
Vibration resistance Compatible with JIS Vibration testing methods for automobile
Type 1 Class A-equivalent
Certifications CE, RoHS, China RoHS

GL800 Input Terminal Unit Specifications

Item Description
Number of input channels 20 (maximum 200 channels by the expansion terminal unit)
Type of input terminal M3 screw type terminal
Method Scan
All channels isolated, balanced input
Measurement range Voltage 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50V, 1-5 V/F.S.
Temperature Thermocouple : K, J, E, T, R, S, B, N, W (WRe5-26)
RTD : Pt100, JPt100, Pt1000 (IEC751)
Humidity 0 to 100 % RH (Voltage 0 to 1 V conversion, when using optional B-530 humidity sensor is used)
Input filter Off, 2, 5, 10, 20, 40 (Moving average)
Measurement accuracy (23°C ± 5°C)
  • When 30minutes
    or more have
    elapsed after
    power was
    switched on
  • Sampling 1s/20ch
  • Filter ON(10)
  • GND connected
Voltage ± 0.1 % of F.S.
Temperature Measurement range Accuracy
Thermocouple R/S 0 ≦ TS ≦ 100 °C ± 5.2 °C
00 < TS ≦ 300 °C ± 3.0 °C
R : 300 < TS ≦ 1600 °C ± (0.05 % of rdg + 2.0 °C)
S : 300 < TS ≦ 1760 °C ± (0.05 % of rdg + 2.0 °C)
B 400 ≦ TS ≦ 600 °C ± 3.5 °C
600 < TS ≦ 1820 °C ± (0.05 % of rdg + 2.0 °C)
K -200 ≦ TS ≦ -100 °C ± (0.05 % of rdg + 2.0 °C)
-100 < TS ≦1370 °C ± (0.05 % of rdg + 1.0 °C)
E -200 ≦ TS ≦ -100 °C ± (0.05 % of rdg + 2.0 °C)
-100 < TS ≦ 800 °C ± (0.05 % of rdg + 1.0 °C)
T -200 ≦ TS ≦ -100 °C ± (0.1 % of rdg + 1.5 °C)
-100 < TS ≦ 400 °C ± (0.1 % of rdg + 0.5 °C)
J -200 ≦ TS ≦ -100 °C ± 2.7 °C
-100 ≦ TS ≦ 100 °C ± 1.7 °C
100 < TS ≦ 1100 °C ± (0.05 % of rdg + 1.0 °C)
N 0 ≦ TS ≦ 1300 °C ± (0.1 % of rdg + 1.0 °C)
M 0 ≦ TS ≦ 2000 °C ± (0.1 % of rdg + 1.5 °C)
* If the reference junction compensation is internal, add ± 0.5 °C to each of the above values.
RTD Pt100 -200 to 850 °C (FS = 1050 °C) ± 1.0 °C
JPt100 -200 to 500 °C (FS = 700 °C) ± 0.8 °C
Pt1000 -200 to 500 °C (FS = 700 °C) ± 0.8 °C
A/D converter 16 bit (out of which 14 are internally acknowledged)
Maximum input voltage 60 Vp-p (Between ±)
60 Vp-p (Between input terminals)
60 Vp-p (Between input terminal and chassis)
Withstand voltage 350 Vp-p (between input terminal and GND) 1 minute
*1
The available number of channels varies according to the sampling interval.
*2
Max. input voltage : 24 V, Input threshold voltage : approx. 2.5 V, Hysteresis approx. 0.5 V (+2.5 to 3 V)
*3
Option
*4
Excluding the battery and AC adapter
*5
Logic/Alarm cable (B-513) is required