Instrument products > midi LOGGER

GL200A

Technical Specs & Details

GL200A main unit specifications

Item Description
Number of input channels Analog : 10 ch, Pulse *1 : 1 ch, Logic *1 : 1 ch
Analog input terminal shape Screw terminal
Sampling interval 10 ms to 1 h (10 ms to 50 ms are for voltage measurement only, there is a limit to the number of channels)
Measurement method Scanning method
Measurement range Voltage 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50 V, 1-5 V / F.S.
Temperature Thermocouple : K, J, T, R, S, B, N, W (WRe5-26)
Humidity 0 to 100 % ( when using optional humidity sensor, 0-1 V scaling conversion is used)
Filter Off, 2, 5, 10, 20, 40 (Moving average with following values isdone)
Measurement accuracy Voltage 0.1 % of F.S.
Temperature Thermo couple Measurement Temperature Range (°C) Measurement Accuracy
R/S 0 ≤ TS ≤ 100
100 < TS ≤ 300
R : 300 < TS ≤ 1600
S : 300 < TS ≤ 1760
Reference contact compensation accuracy : ±0.5°C
±5.2°C
±3.0°C
±(0.05% of rdg +2.0°C)
±(0.05% of rdg +2.0°C)
B 400 ≤ TS ≤ 600
600 < TS ≤ 1820
Reference contact compensation accuracy : ±0.5°C
±3.5°C
±(0.05% of rdg +2.0°C)
K -200 ≤ TS ≤ -100
-100 < TS ≤ 1370
Reference contact compensation accuracy : ±0.5°C
±(0.05% of rdg +2.0°C)
±(0.05% of rdg +1.0°C)
E -200 ≤ TS ≤ -100
-100 ≤ TS ≤ 800
Reference contact compensation accuracy : ±0.5°C
±(0.05% of rdg +2.0°C)
±(0.05% of rdg +1.0°C)
T -200 ≤T≤ -100
-100 < TS ≤ 400
Reference contact compensation accuracy : ±0.5°C
±(0.1% of rdg +1.5°C)
±(0.1% of rdg +0.5°C)
J -200 ≤ TS ≤ -100
-100 < TS ≤ 100
100 < TS ≤ 1100
Reference contact compensation accuracy : ±0.5°C
±2.7°C
±1.7°C
±(0.05% of rdg +1.0°C)
N 0 ≤ TS ≤ 1300
Reference contact compensation accuracy : ±0.5°C
±(0.1% of rdg +1.0°C)
W 0 ≤ TS ≤ 2315
Reference contact compensation accuracy : ±0.5°C
±(0.1% of rdg +1.5°C)
Reference contact compensation Internal/External switching
A/D converter 16 bits (out of which 14 are internally acknowledged)
Trigger Functions Trigger types Start: Data capture starts when a trigger is generated.
Stop: Data capture stops when a trigger is generated.
Trigger conditions Start: Off, Level *2, External, Time
Stop: Off, Level *2, External, Time, Specified period of time
Alarm functions Type Analog, Logic or Pulse; OR logic
Condition*2 Analog: Rising, Falling, Window In, Window Out
Pulse: Rising, Falling, Window In, Window Out
Logic: Rising, Falling
Pulse input mode Count mode Displays a count of the number of pulses for each sampling interval from the start of measurement
Inst mode Counts the number of pulses for each sampling interval. Resets the count value after each sampling interval.
RPM mode Counts the number of pulses per second; enables them to be converted to rpms.
Pulse input range Count mode 50 c, 500 c, 5000 c, 50 kc, 500 kc, 5 Mc, 50 Mc, 500 Mc
Inst mode 50 c, 500 c, 5000 c, 50 kc, 500 kc, 5 Mc, 50 Mc, 500 Mc
RPM mode 50 rpm, 500 rpm, 5000 rpm, 50 krpm, 500 krpm, 5 Mrpm, 50 Mrpm, 500 Mrpm
Alarm output Number of channels 1 ch
Output type Open collector output (100 kΩ pull-up resistance)
Output conditions Level judgment, window judgment, logic pattern judgment, pulse judgment
External trigger input*1 1 ch
PC I/F USB
Data storage functions Measurement data Direct capture to the internal flash memory or USB memory stick
Other Setting parameters and screen copy data can also be saved
Memory devices GL200A internal flash memory (3.5 Mbytes), USB memory stick
Calculation functions Statistical calculation: Average value, Peak value, Maximum value, Minimum value, RMS
Search function Search for analog signal levels or alarm generation points in captured data
Search types Analog signal Rising or falling with respect to the specified level
Alarm Both, Rising, Falling
Scaling conversion function Input (upper and lower limits) and output (upper and lower limits) can be set for each channel
Display unit Size: 3.5-inch TFT color LCD; Display information: Waveforms + digital values, waveforms only, digital values only
Maximum permissible input voltage Between +/- terminals: 60 Vp-p
Between input terminals and casing: 60 Vp-p
Withstand voltage Between each input channel and main unit chassis, and also between each CHs: 1 minute at 350 Vp-p
Operating environment 0 to 40°C, 30 to 80% RH
Power supply AC adapter: 100 to 240 VAC, 50/60 Hz
DC input: 8.5 to 24 VDC *3
Battery pack (option) *3
Power consumption 28 VA or lower (when the AC drive is used)
External dimensions (W x D x H) (approx.) 194 x 122 x 41 mm
Weight (approx.) 480 g (excluding AC adapter and battery)
*1
Maximum input voltage: +24 V, Input threshold voltage: Approx. +2.5 V, Hysteresis: Approx. 0.5 V (+2.5 V to 3 V)
*2
The trigger condition operation is Level trigger. Measurement starts when the condition specified for the start of measurement is satisfied and the trigger is activated.
*3
The DC drive input cable and battery pack are optional.

Control software specifications

Item Description
CPU Pentium 4, 1.7 GHz or more
Memory At least 512MB (1 GB recommended)
Supported OS Windows 2000, Windows XP, Windows Vista
Functions GL200A control, real-time data capture, file format conversion
GL200A setting range Input settings, memory settings, alarm settings, trigger settings
Captured data Real-time transfer to a PC, transfer of data from the GL200A’ s internal memory
Display information Analog waveforms, logic waveforms, pulse waveforms, digital values
Display modes Digital values, waveforms
Monitoring function An email is sent to a specified email address when an alarm is generated
File format conversion Conversion of data between cursors or all data to the CSV format
Direct to Excel Saves sampling data up to 100 ms to an Excel file
Maximum/minimum Displays the maximum, minimum, and current values during measurement